Chapter Summary

Option 048: In the bulleted list delete:

Internal Battery Maintenance 7-11 Where to Find Important Topics

Option 048: In the bulleted list delete:

Disconnecting the Internal Battery 7-13

Removing and Replacing the Internal Battery 7-13

Using the Battery Disconnect Switch 7-12

Preparation for Use

Power Requirements

Option 048: 2nd paragraph, 1st and 2nd sentences are changed:

Operating power may also be supplied as external dc voltages between 40 and 58 Vdc. Maximum dc-power consumption is 100 W.

Bench Operation

Option 048: 1st WARNING sentence is changed-

The 5071A weighs 28 kg (61 lbs).

Operating the 5071A From External Dc Power

Option 048: 1st paragraph, 1st sentence is :

In addition to ac power capability, the 5071A can also operate from an external dc source with a voltage range of ± 40 to 58 Vdc.

Option 048: Note content following the first paragraph is changed:

The 5071A runs from dc power when both ac and dc power are available. The 5071A will not run from ac power unless the dc power source fails or is disconnected.