Assembly-Level Service

Service

1.4.4.2.1A1C4 Voltage Test Procedure

Option 048: The QUESTION, is changed to:

Is the voltage between 45 and 51 Vdc?

1.4.4.3LF1/T1 Transformer Fault Verification Option 048: Step 2 text is changed to:

Verify presence of 33 - 47 Vrms.

The QUESTION: is changed to:

Is the voltage between 33 - 47 Vrms?

The T1 part number within If No response text to the QUESTION., is changed to:

(9100-5134)

Advisory Messages (Subsection 7)

1.7.1Advisory Message Interpretation

Any TCO Option: Advisory Message: “SAW Osc unlocked” is deleted,

Functional-Group Diagnostic Trees

(Diagnostic Section 2)

Option 048:

Option 048:

Option 048:

1 Power-Supply Diagnostic Tree: 2.1.1.1.0

The bulleted item “B1 24V Internal-Standby Battery” on page 2-45 is deleted.

Power-Supply Diagnostic Tree

(Functional Group Subsection 1)

The bulleted item “B1 24V Internal-Standby Battery” on page 2-46 is deleted.

2.1.1.1.1 A12 Power Input Test

Step 2 4 text is changed to:

Verify that the voltage present is between 45 and 51 Vdc with no more than 1 Vrms ripple.

The second sentence of the If No response to the first QUESTION: is changed:

b)Verify that the voltage present is between 45 and 51 Vdc with no more than 1 Vrms ripple.