Operating and Programming

Getting Started

Getting Started

This section describes the additions, changes, and deletions to the information that appears in chapter 1 of the 5071A Operating and Programming manual.

CHAPTER GUIDE

Where to Find Important Topics

Option 048: In the bulleted list delete:

Powering Down the 5071A 1-8

The 5071A at a Glance

Connectors

Any TCO Option: The paragraph immediately following item 8. Sync Input is changed-

(The electrical characteristics of these connectors are the same as the rear-panel 1 pps Output and Sync Input. See “Rear-Panel Features” on page 4 and 5 for more information.) The rear-panel 1 pps outputs and sync input connectors are absent. The front-panel 1 pps output and sync input signal characteristics are however accurately described by items 5 and 6 of Table 1-1 “Connector Descriptions” in the Operating and Programming manual.

Any TCO Option: Option 048:

Rear-Panel Connections

Table 1-1 Connector Descriptions items in the Operating and Programming manual are changed:

Items 5 and 6 are absent from the rear-panel.

Item 9 is absent.

Item 10 Connector Type is a 5-screw terminal strip.

Item 10 Signal Characteristics are 40-58 Vdc on the rear-panel terminal strip dc-power inputs 1 and/or 2 with GND as chassis ground.