Site Preparation
Site Preparation Guide 23
Electrical Requirements
You are responsible for providing appropriate electrical power and power
outlets for all of the components in your 5975 Series MSD system. Power
considerations include:
Voltage ranges of major components
Power configurations
Power requirements
Power plug s and cord s

Voltage ranges of major components

The 5975 Series MSD includes a full-range power supply that can operate
without reconfiguration on either of two wide ranges of single-phase
alternating current (AC) electrical power:
120-127 VAC 60 Hz (typical for North America)
200-240 VAC, 50/60 Hz (typical for Europe and Japan)
The foreline pump also draws its power from the MSD. However, a different
foreline pump is supplied depending on the voltage range on which the MSD
will be operating. The pump is supplied according to the standard voltage in
the country from which the order originates. For example, if an order
originates from an Agilent Technologies sales office in Germany, the foreline
pump supplied will be configured to operate on the standard voltage and
frequency of electrical power in Germany.
NOTE The MSD does not support 100 VAC input power. In countries that have only 100 VAC
power, a step-up transformer must be used to provide the appropriate voltage to the MSD.
CAUTION If an instrument is being or dered from one location but is to be installed in another
location with different electrical power characteristics, this must be noted on the order.
A special note must also be made if the electrical power at the site is different from the
standard electrical power in that country.