28 Site Preparation Guide
Site Preparation
Air Conditioning and Environmental Requirements
Air conditioning considerations includ e temperature, humidity, airborne dust,
and exhaust venting. Each of these is considered in more detail in the
following m aterial.

Temperature, humidity, and altitude

The 5975 Series MSD is specified for operation under the following conditions:
Operation requires constant temperature
(variations <2 °C/h)
Operation and storage require a noncondensing, noncorrosive atmosphere
The temperature and humidity limits in Table 5
Maximum altitude for operation: 2300 m (7500 ft)
Environmental control systems must maintain thes e temperature and
humidity ranges.
The MSD is rated for 1200 Watts (4000 BTU/h). The GC and the data system
also contribute significantly to the cooling load although the exact amounts
depend on their configurations. Additional allowances should be made for
other heat sources such as heat from other equipment, heat from adjac ent
rooms, and heat from laboratory personnel.
Tabl e 5 Temperature and humidity limits
Temperature Humidity (relative)
Operation 15 °C to 35 °C (59 °F to 95 °F) 40% to 80%
Storage –20 °C to 70 °C (–4 °F to 158 °F) 0% to 95%