32 Site Preparation Guide
Site Preparation
Regulators, tubing, and fittings
You must supply appropriate regulators for your carrier and reagent gas.
Regulators must be dual-stage.
Regulators must have stainless-steel (not elastomeric) diaphragms.
They must supply gas at the pressure specified.
They must have one outlet with 1/4-inch Swagelok fittings.
See the Agilent Technologies Chemical Analysis Columns and Supplies
Catalog or visit the Agilent Technologies Chemical Analysis website at
http://www.agilent.com/chem for dual-stage regulators available from Agilent
Technol ogies.
You must supply Swagelok or Swagelok-compatible fittings, ferrules, and
connectors for the 1/8-inch tubing.
5975 Series Site Prep.book Page 32 Thursday, January 31, 2008 9:47 AM