Site Preparation

Carrier and Reagent Gas Requirements

Table 6 lists the specifications for the carrier and reagent gases. Pressures and flows are as measured at the bulkhead fittings on the back of the instrument, not at the gas supply. We recommend that additional shut-off valves be installed near the instruments.

Table 6

Carrier and reagent gas requirements

 

 

 

 

 

 

 

 

 

Purity

Typical pressure range,

Typical flow, mL/min

 

 

 

kPa (psi)

 

 

 

 

 

Helium (required)

99.9995% or better and

345 to 550 (50 to 80)

20 to 50 (column and split flow)

 

 

hydrocarbon-free

 

 

 

 

 

 

Methane reagent gas

Research or SFC grade,

103 to 172 (15 to 25)

1 to 2

(required for CI operation)

99.999%

 

 

 

 

 

 

Isobutane reagent gas

Instrument grade,

103 to 172 (15 to 25)

1 to 2

(optional)

 

99.99%

 

 

 

 

 

 

Ammonia reagent gas

Research or SFC grade,

34 to 55 (5 to 8)

1 to 2

(optional)

 

99.9995%

 

 

 

 

 

 

Carbon dioxide reagent gas

SFC Grade,

103 to 138 (15 to 20)

1 to 2

(optional)

 

99.995%

 

 

 

 

 

 

 

WA R N I N G

Methane is flammable and ammonia is corrosive and toxic. Read and follow the

safety information that comes with the gases.

 