Site Preparation

 

 

 

 

Table 4

Power requirements (continued)

 

 

 

 

 

 

 

 

 

Product

 

Line voltage

Maximum continuous

Supply

Outlets

 

 

 

power consumption

circuit rating

required

 

 

 

 

 

ChemStation computer

100-120 or 220-240 VAC+5/-10%,

1000 VA

15 A

3 to 5

 

 

50/60 Hz ± 5%

 

 

 

 

 

 

 

 

 

*The MSD operates on either voltage range. The foreline and turbo pumps are supplied for the voltage range suitable to the country where the order originates

† Americas 120 V requires 20 A dedicated line. Americas 240 V requires 15 A dedicated line.

Option 003, fast oven, is a unit with operating range 198 to 231 V. Most laboratories have 4-wire service resulting in 208 V at the wall receptacle. It is important to measure the line voltage at the receptacle for the GC.

** Requires an isolated ground and dedicated outlet.

Power plugs and cord

 

The MSD is supplied with a power cord and plug appropriate for the country

 

from which the order originates. For example, if an order originates from a

 

Agilent Technologies sales office in Germany, the power cord and plug

 

supplied will be compatible with the standard voltage and outlet configuration

 

in Germany. See “Power Cords" on page 41 for illustrations of the power cords

 

available.

 

Data system components also include power cords with plugs appropriate for

 

the country where the order was placed.

 

Power cord lengths for the MSD and the data system components and

 

accessories are approximately 2.3 m (7.5 ft).

 

If an instrument is being ordered from one location but is to be installed in another

C A U TI O N

location with different electrical power characteristics, this must be noted on the order.

 

 

A special note must also be made if the electrical power at the site is different from the

 

standard electrical power in that country.

 

 

 

Make sure the power cords supplied with the MSD are appropriate for your country

WA R N I N G

and site before installing the instrument. See “Power Cords” on page 41.

 

 

 

26

Site Preparation Guide

Page 26
Image 26
Agilent Technologies 5975 manual Power plugs and cord, Supply Outlets Power consumption Circuit rating Required