2 - Verification and Performance Tests

2 - Verification and Performance Tests

Resistor

1 ohm, 50 W

Ohmite L50J1R0

(substitute for electronic

3 ohm, 100 W (Agilent 66332A/6632B)

Ohmite RLS5R0 (adjustable)

load if load is too noisy

24 ohm, 100 W (Agilent 6633B)

Ohmite RLS25R (adjustable)

for CC PARD test)

99 ohm, 100 W (Agilent 6634B)

Ohmite RLS100 (adjustable)

 

1k ohm, 5%, 3W (all models)

Agilent 0813-0001

Oscilloscope

Sensitivity: 1 mV

Agilent 54504A or equivalent

 

Bandwidth Limit: 20 MHz

 

 

Probe: 1:1 with RF tip

 

RMS Voltmeter

True RMS

Agilent 3400B or equivalent

 

Bandwidth: 20 MHz

 

 

Sensitivity: 100 μV

 

Variable-Voltage

Adjustable to highest rated input voltage range.

 

Transformer

Power: 500 VA

 

Measurement Techniques

Test Setup

Most tests are performed at the rear terminals of the supply as shown in Figure 2-1a. Measure the dc voltage directly at the +S and -S terminals.

+ 240 VDC MAX

Measurement Techniques -

+S + - - S

Test Setup - + 240 VDC MAX

+S + - -S

DVM, Scope, or

+

 

RMS voltmeter

 

(for CV tests)

-

 

 

DVM or

+

 

RMS voltmeter

 

(for CC tests)

-

 

 

 

Current

monitor

+-

Electronic

Load

(see note)

+

DC

Ammeter

-

b.

+

DC

Ammeter

-

Load resistor 400 ohm

-+ 240 VDC MAX

+S + - -S

Load

resistor 1 k ohm

 

Note: Use dc supply with same polarity

 

connections for - CC tests.

a.

Replace load with appropriate

 

resistor for CC noise test.

c.

+-

External

DC supply

Figure 2-1. Test Setup