3 - Troubleshooting

Manual Fan Speed Control

Under some circumstances such as testing acoustical devices where the fan noise would interfere with the test, it would be advantageous to reduce the fan speed. If the test requires a very light load, the ambient temperature is low and the duration of the test is short, the fan speed may be temporarily reduced. The turn-on default is "Automatic" so this procedure must be performed, as needed, every time the line voltage is turned on. To manually control the fan speed:

a.Simultaneously depress the "0" and "9” keys. EEINIT <model> will be displayed.

b.Using the Up/Down annunciator keys select FAN:MODE<AUTO.>.

c.Using the Up/Down arrows select FAN:MODE <MAN>

d.Press "Enter"

e.Simultaneously depress the "0" and "9" keys. EEINIT <model> will be displayed.

f.Using the Up/Down annunciator keys select FAN:SPEED <data>

g.Press "Enter Number".

h.Enter the desired speed (numeric entry range is 0 to 100%)

i.Press "Enter"

Disabling Protection Features

Except for overvoltage protection, the power supply's protection features may be disabled. This is not recommended as a normal operating condition but is helpful under some circumstances such as troubleshooting. The turn-on default is "NO-PROTECT OFF" (protection enabled) so this procedure must be performed, as needed, every time the line voltage is turned on. To disable the protection:

a.Simultaneously depress the "0" and "9" keys. EEINIT <model> will be displayed.

b.Using the Up/Down annunciator keys select NO-PROTECT <OFF>.

c.Using the Up/Down arrows select NO-PROTECT <ON>.

d.Press "Enter"

40

Page 40
Image 40
Agilent Technologies 6634B, 66332A, 6633B, 6632B Manual Fan Speed Control, Disabling Protection Features, Troubleshooting