Agilent Technologies 66332A, 6634B, 6633B, 6632B service manual Verification and Performance Tests

Models: 6632B 6633B 66332A 6634B

1 83
Download 83 pages 60.43 Kb
Page 21
Image 21

Verification and Performance Tests - 2

Model Agilent 6633B

Report No _______________

 

Date __________________

Test Description

 

Minimum

 

 

Results*

Maximum

 

 

Specs.

 

 

 

Specs.

Constant Voltage Tests

 

 

 

 

 

 

Voltage Programming and Readback

 

 

 

 

 

 

Low Voltage (0V) Vout

 

− 20 mV

 

__________

+ 20 mV

Front Panel Display Readback

 

Vout − 6 mV

 

__________

Vout + 6 mV

High Voltage (Full Scale) Vout

 

49.955 V

 

__________

50.045 V

Front Panel Display Readback

 

Vout − 21 mV

 

__________

Vout + 21 mV

Load Effect

 

− 4 mV

 

__________

+ 4 mV

Source Effect

 

− 1.0 mV

 

__________

+ 1.0 mV

PARD (Ripple and Noise)

 

 

 

 

 

 

Peak-to-Peak

 

N/A

 

__________

+ 3 mV

RMS

 

N/A

 

__________

+ 0.5 mV

Transient Response

 

 

 

 

 

 

Voltage in 100 μs

 

Vout − 50 mV

 

__________

Vout + 50 mV

Constant Current Tests

 

 

 

 

 

 

Current Programming and Readback(High Range)

 

 

 

 

 

CC Programming Accuracy @ 0A

 

− 1.53 mA

 

__________

+ 1.53 mA

Current Readback Accuracy @ 20mA (High Range)

Iout − 0.29 mA

 

__________

Iout + 0.29 mA

CC Programming @ Full Scale

 

1.998 A

 

__________

2.002 A

Current Readback Accuracy @ Full Scale

 

Iout − 4.25 mA

 

__________

Iout + 4.25 mA

Current Sink Readback

 

Isink − 4.85mA

 

__________

Isink + 4.85mA

20 mA Range Current Readback

 

 

 

 

 

 

Readback Accuracy @ 0 A

 

− 2.5 μA

 

__________

+ 2.5 μA

Readback Accuracy @ + 20 mA

 

Iout − 22.5 μA

 

__________

Iout + 22.5 μA

Readback Accuracy @ − 20 mA

 

Iout − 22.5 μA

 

__________

Iout + 22.5 μA

PARD (Current Ripple and Noise)

 

 

 

 

 

 

RMS

 

N/A

 

__________

+ 2.0 mA

Load Effect

 

− 1.0 mA

 

__________

+ 1.0 mA

Source Effect

 

− 0.25 mA

 

__________

+ 0.25 mA

* Enter your test results in this column

 

 

 

 

 

 

21

Page 21
Image 21
Agilent Technologies 66332A, 6634B, 6633B, 6632B service manual Verification and Performance Tests