Agilent Technologies 6633B, 6634B, 66332A Constant Voltage CV Tests, CV Setup, CV Load Effect

Models: 6632B 6633B 66332A 6634B

1 83
Download 83 pages 60.43 Kb
Page 14
Image 14
Table 2-2. Programming Ratings

2 - Verification and Performance Tests

Table 2-2. Programming Ratings

Model

Voltage Rating

Full Scale Rating

Current Rating

Full Scale Rating

Agilent

20 V

20.020 V

5 A

5.0045 A

66332A/6632B

50 V

50.045 V

2 A

2.002 A

Agilent 6633B

100 V

100.1 V

1 A

1.001 A

Agilent 6634B

 

 

 

 

Constant Voltage (CV) Tests

CV Setup

If more than one meter or if a meter and an oscilloscope are used, connect each to the terminals by a separate pair of leads to avoid mutual coupling effects. For constant voltage dc tests, connect only to +S and -S, since the unit regulates the output voltage that appears between +S and -S, and not between the (+) and (-) output terminals. Use coaxial cable or shielded two-wire cable to avoid noise pickup on the test leads.

Voltage Programming and Readback Accuracy

This test verifies that the voltage programming, GPIB readback and front panel display functions are within specifications. Note that the values read back over the GPIB should be identical to those displayed on the front panel.

a.Turn off the supply and connect a digital voltmeter between the +S and the -S terminals as shown in Figure 2-1a.

b.Turn on the supply and program the supply to zero volts and the maximum programmable current with the load off.

c.Record the output voltage readings on the digital voltmeter (DVM) and the front panel display. The readings should be within the limits specified in the performance test record chart for the appropriate model under CV PROGRAMMING @ 0 VOLTS. Also, note that the CV annunciator is on. The output current reading should be approximately zero.

d.Program the output voltage to full-scale.

e.Record the output voltage readings on the DVM and the front panel display. The readings should be within the limits specified in the performance test record chart for the appropriate model under CV PROGRAMMING @

FULL SCALE.

CV Load Effect

This test measures the change in output voltage resulting from a change in output current from full load to no load.

a.Turn off the supply and connect the output as shown in Figure 2-1a with the DVM connected between the +S and -S terminals.

b.Turn on the supply and program the current to the maximum programmable value and the voltage to the full- scale value.

c.Adjust the load for the full-scale current as indicated on the front panel display. The CV annunciator on the front panel must be on. If it is not, adjust the load so that the output current drops slightly.

d.Record the output voltage reading on the DVM connected to +S and -S.