Agilent Technologies 6634B, 66332A, 6633B, 6632B Troubleshooting, 1Sheet 12. High Output Voltage

Models: 6632B 6633B 66332A 6634B

1 83
Download 83 pages 60.43 Kb
Page 36
Image 36
Figure 3-1 Sheet 12. High Output Voltage

3 - Troubleshooting

Disable the OV capability by

shorting R351. After the

protection is disabled, program

the output voltage to zero,

current to full scale and Output ON. If the unit is in "Protect" mode, Press Protect Clear. The output should now go high and not trip the OV.

*V_mon should be approximately 6632B or 66332A Vout/4.25

6633B Vout/10.52

6634B Vout/21

Is the CV

annunciator on

?

Yes

Measure voltage at the base of Q303 with respect to its' emitter

Voltage <0.6V

?

Yes

 

Vmon,

 

 

 

Check Voltage Monitor

No

U315B-7,

 

No

 

Amplifier, U315B

 

 

OK ?*

 

 

 

circuit

 

 

 

 

 

 

Yes

CV_Prog,

R401 No Manual background A2 Interface Board

~0V

?

Yes

Check Voltage Control

U315A, circuit

No

Troubleshoot Voltage

Gain Stage

 

Troubleshoot Output

Stage

Figure 3-1 Sheet 12. High Output Voltage

36

Page 36
Image 36
Agilent Technologies 6634B, 66332A, 6633B, 6632B service manual Troubleshooting, 1Sheet 12. High Output Voltage