Figure 3-1 Sheet 4. OV at Turn-On

3 - Troubleshooting

Connect a DC coupled scope set to 1mS/20V/ div across the output and turn on the supply while observing the scope for a momentary pulse greater than the supply rating

Does the supply

 

 

 

 

Go to “Troubleshooting

 

Yes

 

 

High Output Voltage”

overshoot?

 

 

 

 

 

 

(sheet 7)

 

 

 

 

 

 

 

 

 

 

 

No

Disable the OV circuit as described in paragragh "Disabling Protection Features"

Output @ zero

 

 

 

No

volts?

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Yes

 

 

 

4V @ A1R350-2? NoManual background A2 or W7 Defective

Yes

Check C336, R356,

+.3V @ U306B-7? NoManual background R351, R349(1-2) and

U306B

Yes

Check C335, R354,

+4V @ U306-8?No R350(1-2) and U306B

Yes

Go to sheet 5

Figure 3-1 Sheet 4. OV at Turn-On