Line Voltage Wiring

3 - Troubleshooting

Line Voltage Wiring

Figure 3-3 illustrates the primary wiring configuration of the power transformer for various ac line voltages. Use long nose pliers to disconnect the wires going to the transformer terminals.

NOTE: Install the correct fuse when changing the ac line voltage from a previous setting: for 110/120 Vac: 4 AM, Agilent p/n 2110-0055;

for 220/230 Vac: 2 AM, Agilent p/n 2110-0002

 

grey

grey

white/red/grey

 

 

 

 

 

white/red/grey

spare

 

jumper

jumper

 

 

grey

white/red/grey

grey

white/red/grey

jumper

jumper

white/blue

white

orange

white/grey

white/green

 

yellow

grey

 

blue

white/red

 

white/red

 

Figure 3-3. Transformer Wiring