Agilent Technologies 669xA, 665xA, 664xA, 667xA, 668xA manual Connecting Supplies in Auto-Parallel

Models: 668xA 669xA 667xA 664xA 665xA

1 142
Download 142 pages 24.05 Kb
Page 73
Image 73
Load Connection
B Connect for optional remote sensing

 

 

 

Loads

Load Connection

Analog Connector

A Connect for remote sensing (optional)

B

Connect for local sensing (default)

Figure 4-4d. Series 667xA Multiple Load Connection (Remote Sensing Optional)

Connecting Supplies in Auto-Parallel

Auto-Parallel Wiring (Figure 4-4e).Figure 4-4e illustrates how power supplies can be connected in auto-parallel for increased current output. You can connect up to five supplies of the same model.

Use load leads of a sufficient wire size so that the absolute voltage difference between the + output terminal of the "master" supply and the + output terminal of the first "slave" supply is kept under 2 V at rated current. This also applies to the voltage difference between the + output terminals of the first and second slave supplies. If remote sensing is required, connect the load to the remote sense terminals of the master supply, as shown by the dashed lines in Figure 4-4e.

Analog ConnectorSlave SupplyMaster Supply

Program only the master. Set slave output and OVP voltage slightly higher than the master to ensure that slave stays in CC mode

Load

A Only local sensing permitted

Figure 4-4e. Series 667xA Auto-Parallel Connection (Remote Sensing Optional)

User Connections 73