Agilent Technologies 669xA, 665xA Maximum AC Line Current Ratings Vac nominal, Load Regulation

Models: 668xA 669xA 667xA 664xA 665xA

1 142
Download 142 pages 24.05 Kb
Page 18
Image 18

Table 1-lb. Supplemental Characteristics for Series 664xA (continued)1

Parameter

 

 

Agilent Model Number

 

 

6641A

 

6642A

6643A

6644A

6645A

Maximum AC Line Current Ratings

 

 

 

 

 

 

100 Vac nominal:

 

 

 

4.4 A rms

 

 

120 Vac nominal:

 

 

 

3.8 A rms

 

 

220 Vac nominal:

 

 

 

2.2 A rms

 

 

230 Vac nominal:

 

 

 

2.1 A rms

 

 

240 Vac nominal:

 

 

 

2.0 A rms

 

 

Maximum Reverse Bias Current:

With AC input power applied and the dc output reverse

 

biased by an external dc source, the supply will continuously

 

withstand without damage a current equal to its output

 

current rating (see Table 1- 1a).

 

 

Remote Sensing Capability

 

 

 

 

 

 

Voltage Drop Per Lead:

Up to 1/2 of rated output voltage.

 

 

Load Regulation:

Add 3 mV to spec (see Table l-la) for each l-volt change in

 

the + output lead due to load current changes.

 

Load Voltage:

Subtract voltage drop in load leads from specified output

 

voltage rating.

 

 

 

 

Command Processing Time (Average time for output voltage to change after receipt of digital data when the supply is connected directly to the GPIB Bus):

 

 

 

 

 

 

20 ms

 

 

 

Downprogrammer Current Capability (± 15%):

 

5.8 A

 

2.5 A

 

1.5 A

 

0.9 A

 

0.75 A

 

 

 

 

 

 

Output Voltage Programming Response Time

 

 

 

 

 

 

 

 

 

 

Rise/Fall Time (time for output to change from 90 % to 10% or from 10% to 90% of its total excursion):

 

 

Settling Time (time for output change to settle within

 

<15 ms

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1 LSB (0.025% x rated voltage) of its final value):

 

 

 

 

<60 ms

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Monotonicity:

 

Output is monotonic over entire rated voltage, current, and

 

 

temperature range.

 

 

 

 

 

 

Auto-Parallel Configuration:

 

 

 

Up to 3 identical models

 

 

 

 

 

 

 

 

 

 

 

 

 

Analog Programming (IP & VP)

 

 

 

 

 

 

 

 

 

 

Input Signal:*

 

 

 

 

 

0 to -5 V

 

 

 

Input Impedance:

 

 

 

 

10 kΩ, nominal

 

 

 

*Signal source must be isolated.

 

 

 

 

 

 

 

 

 

 

Current Monitor Output (+IM):

 

0 to -5 V represents zero to full-scale current output

 

 

 

 

 

 

 

 

 

 

 

 

 

Savable States

 

 

 

 

 

 

 

 

 

 

Nonvolatile Memory Locations:

 

 

 

 

5 ( 0 through 4)

 

 

 

Nonvolatile Memory Write Cycles:

 

 

 

 

40,000, typical

 

 

 

Prestored State (factory default):

 

 

 

 

 

Location 0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Notes: lFor Performance Specifications, see Table l-la.

18 General Information

Page 18