Agilent Technologies 668xA, 665xA, 664xA Programming Voltage, Programming Overvoltage Protection

Models: 668xA 669xA 667xA 664xA 665xA

1 142
Download 142 pages 24.05 Kb
Page 89
Image 89
. The display will change from meter mode to indicate VOLTS.
. If you discover a mistake before pressing , erase the incorrect value with the backspace

Figure 5-2. Typical Power Supply Operating Curve

Programming Voltage

To program the output for 4.5 volts, proceed as follows:

Press

Press

key .

The display will return to the meter mode and indicate 0.000 volts.

Press to enable the output (Dis annunciator turns off). The VOLTS display will indicate 4.500 volts.

Note

The power supply must be programmed for a minimal current in order to increase the output voltage

 

beyond zero. Normally, there is sufficient idle current to do this. If the power supply does not respond or

 

the CC annunciator turns on, go to “Programming Current” and set the current to a small value.

Now raise the voltage by pressing .Note that the voltage increases by a specific increment (depending on the voltage programming resolution) each time you press the key and increases rapidly as you hold down the key. To lower

the voltage, press .

Try raising and lowering the voltage by rotating the Voltage control clockwise and then counterclockwise. Note how the output responds as compared to using the Entry keys.

Try to program a voltage greater than the VMAX for your supply (see "Supplemental Characteristics" for your particular model in Chapter 1). Note that the display shows OUT OF RANGE.

Programming Overvoltage Protection

Overvoltage protection guards the load against voltages that reach a specified value above the programmed output voltage.

Setting the OVP Level

Assuming that you have programmed the power supply for 4.5 volts, you can set the OVP level to 4.8 volts as follows:

Press . The display will change from meter mode to indicate 0V, followed by the present OVP value.

Press .

The display will return to the meter mode and indicate the output (4.500 volts).

Press again. The display will now indicate 0V 4 . 800.

Press to return to the meter mode.

Front Panel Operation 89