94 Installation and User’s Guide
3Verifying System Functionality
The system functional verification test is a standard IC-CAP measurement
and simulation procedure. The system performs DC and S- parameter
extraction, optimization, and simulation using a fixed 10 dB attenuator as
the device under test (DUT).
The procedure provided in “Performing the System Functional Verification
Test” confirms:
•the IC-CAP software can c ommunicate with and control the system
instrumentation via GPIB (through the LAN/GPIB gateway, if so
configured)
•the system can make measurements and display the results
•the IC-CAP software can simu late data
•the IC-CAP software can converge the simulated data with the
extracted (measured) data
The system applies forward and reverse current to the attenuator a nd
monitors voltage at the attenuator’s input and output. IC-CAP then uses
the measured data to model the device- intrinsic resistances and
transmission line delay.
Required Tools
•Agilent 85225F performance modeling system
•A system controller*
•Agilent 85190A IC-CAP sof tware
•Test port cables
•Agilent 8490D 10 dB fixed RF attenuator†
•BNC tee (2)‡
•Agilent 11900A, 2.4 mm male- to-male adapter
•Agilent 11900B, 2.4 mm female- to-female adapter
•Agilent 11900C, 2.4 mm female- to-male adapter
•Agilent 85056A 2.4 mm precision calibration kit, or
•Agilent 85056D 2.4 mm economy calibration kit
* For requirements, see “The S ystem Controller" on page 59 .
† These components are supplied as part of the system.