Agilent 85225F
 Manual Part Number
 Safety and Regulatory Information
Understood and met
 Table a Safety Symbols
SymbolsDefinition
 Table B Instrument Markings
Marking Definition
 Do not Operate in AN Explosive ATMOSPHERE. Do not operate
Must not remove instrument covers. Component replacement
Do not Remove the Instrument COVER. Operating personnel
 Mains power
 Fuses and breakers
For protection against fire hazard, use only fuses with
 Compliance with German Noise Requirements
Acoustic Noise Emission/Geraeuschemission
 This Guide
Introducing the Agilent 85225F Performance Modeling System
 CV Subsystem Functional Verification Tests
 Interface
Interface Examples in Body Text
Tables
Click Enter
 Contents
 Electrical Requirements
To prepare the installation site Environmental Requirements
To unpack the shipment crate containing the rack cabinet
 101
 If you encounter a problem
Bias Network Schematic
 CV, RF, and DC Measurement System Configuration
Performance Modeling System Configuration Overview
RF and DC Measurement System Configuration
Performance Characteristics and Specifications
 Performance Modeling System Configuration Overview
 RF and DC Measurement System Configuration
System Block Diagram
 RF Subsystem
DC Subsystem
 Bias Networks
 Component Integration
 System Components
 Front Panel System Connections, with Agilent 4156C
Number Type Instrument Labeled
 Front Panel Connections with Agilent 4156C
 16493L Triaxial Triax BNC E5260A/70B
16494A Triaxial cable Triax BNC E5260A/70B
Option Gndu cable
 Front Panel Wiring Diagram with Agilent E5260A or E5270B
 10833C Gpib cable
Rear Panel System Connections with Agilent 4156C
10833D Gpib cable
Controller
 Rear Panel Wiring Diagram with Agilent 4156C
 Rear Panel System Connections with Agilent E5260A or E5270B
10833A Gpib cable
 Rear Panel Wiring with Agilent E5260A or E5270B
 Introducing the Agilent 85225F Performance Modeling System
 CV, RF, and DC Measurement System Configuration
 CV Subsystem
 System Components
 Front Panel System Connections
 Front Panel Wiring Diagram
 Rear Panel System Connections
 Rear Panel Wiring Diagram
 DC/RF Cabling Diagram DC and RF Configuration
 DC/RF Cabling Diagram Parametric Configuration
 Low Leakage Switch Mainframe
 Installation and User’s Guide
 Introducing the Agilent 85225F Performance Modeling System
 Noise, CV, RF, and DC Measurement System Configuration
 1/f Noise Subsystem
 F Noise Measurement Block Diagram
 Customer supplied, not included with system
 System Components
 Option Gndu cable 85133F Flexible test E8364B
16494A Triaxial cable Triax BNC 41501B
16493L Triaxial Triax BNC 41501B
Test cable Or probe Station 8120-1839 Coaxial cable
 Front Panel Wiring Diagram
 16493L Gndu cable Triax BNC 41501B
Option 10833A Gpib cable
 Rear Panel Wiring Diagram
 DC/RF Cabling Diagram DC and RF Configuration
 DC/RF Cabling Diagram Parametric Configuration
 Gpib Addresses
Instrument Control Interface
Component Gpib Address
 LAN/GPIB Gateway
 Rear Panel Wiring Diagram for LAN/GPIB Gateway
 Unix Workstation Requirements
System Controller
Personal Computer Requirements
Parameter Requirement
 Rack Cabinet
 Supplemental System Characteristics
Performance Characteristics and Specifications
Performance Modeling System Performance Specifications
Characteristic Value
 Bias Network Characteristics
RF Subsystem Performance Specifications
DC Subsystem Specifications
Station or test fixture characteristics
 To switch on power to the system
To ensure your safety while using the system
Precautions for Avoiding Electrostatic Discharge
Starting on page 91
 Pollution Degree 2 per IEC 61010-1 and 664 respectively
To prepare the installation site
To prepare the installation site
Environmental Requirements
 To receive the system
To receive the system
Step Action
 To unpack the shipment crate containing the rack cabinet
To unpack the shipment crate containing the rack cabinet
Folded position
 Hinged slat
Step Action Unload the system
Wood screws through the ramp
Into the ramp ledge
 To verify the shipment
To verify the shipment
 Replaceable Parts
Part or Model Number Description
 Option AY6 Four-input channel configuration
Dynamic signal analyzer
Option AY2 Two-input channel configuration
Option Triaxial cable, 1.5 meter
 User’s Guide
 Replaceable Parts
 Replaceable Parts
 Prepare to install the work
To install the work surface
To install the work surface
Lock feet prevent the cabinet from
 To install the work surface
 To ensure your safety while using the system
Ground
 Measurements Measurement is in progress
Following section Precautions for
Floating-ground Disconnected and a floating-ground
Fuses with incorrect
 To connect an external ground to the circuit common
Precautions for Performing Floating-Ground Measurements
To perform floating-ground measurements
Circuit common is not connected through the bias networks
 Precautions for Avoiding Electrostatic Discharge
To take precautions against electrostatic discharge
 To connect the bias networks
To connect the bias networks
Agilent 4156C Systems
For Agilent 4156C systems, refer to the following figure
 Bias network Through the upper feedthrough panel
Connector to the port
Bias network open
Connect the bias networks to the device under test
 HRSMU3 Force connector to the port 2 bias network
Agilent 4156C Systems with Agilent 41501B Expander Box
 2.4 mm female-to-male adapter
Port cable to the 2.4 mm female-to-male
 Agilent E5260A/70B Systems
 Connect the triaxial Connect the triaxial cable from
Cables from E5260A/70B medium power SMU to E5281A
High power SMU is the Agilent
Port 2 bias network is
 Connect the bias
Networks to the device
 To switch on power to the system
 Switch on the computer line power
Vector network analyzer line power
Line power
To set the address to 19 and press Enter
 Choose Tools Hardware Setup... .†
To configure the LAN/GPIB gateway
Add Interface
 Installing the System
 Understanding the System Functional Verification Test
To choose a verification process
Performing the System Functional Verification Test
 To choose a system verification process
Performance specifications
To choose a verification process
Situation Action
 Verifying System Functionality
 Understanding the System Functional Verification Test
Required Tools
 Performing the System Functional Verification Test
To perform the system functional verification test
 PC version
 Examples/modelfiles
Choose File Examples
Modelfiles/misc
 Select Calibration Wizard
 Macros tab
 Select Calibrate
 If you encounter a problem
 102
 Servicing the System
 To troubleshoot the system
 Remove the four screws 2 screws
To remove or replace a system component
To remove a system component
Frame nuts
 To order replacement parts
 To receive additional assistance
Contacting Agilent Technologies
 To package the system for transport
To package the system for transport
 Installation and User’s Guide 109
 110
 Performing a Coaxial System Measurement Calibration
To enhance measurement accuracy
Understanding System Measurement Calibration
Periodic System Component Calibration
 To enhance measurement accuracy
 On the network analyzer
Always perform a Prior to making device measurements
Measurement calibration
For more information
 Understanding System Measurement Calibration
 To perform the system measurement calibration
Connector types Connectors dialog box, select
Performing a Coaxial System Measurement Calibration
Networks
 Networks RF/DC OUTPUT. connectors Click Measure
Adapters using Click Measure Standard open, short
Load, and through Short to the Port 1 bias network RF/DC
 Periodic System Component Calibration
 118
 Performing the DC Subsystem Functional Verification Test
Understanding the DC Subsystem Functional Verification Test
 Understanding the DC Subsystem Functional Verification Test
For Agilent E5260A/70B DC subsystems
 Performing the DC Subsystem Functional Verification Test
 122
 Understanding the RF Subsystem Functional Verification Test
 Understanding the RF Subsystem Functional Verification Test
 To perform the RF subsystem functional verification test
Operator’s Check Click Configure
Performing the RF Subsystem Functional Verification Test
Automatic
 126
 Installation and User’s Guide 127
 128
 CV Subsystem Functional Verification
 Understanding the CV Subsystem Functional Verification Test
 Performing the CV Subsystem Functional Verification Test
Manual
 132
 Noise Subsystem Functional
 F Noise Subsystem Functional Verification Test
 If the instrument self test fails, consult Chapter
 136
 Features
Understanding the Bias Networks
 Features
 11612V Option K11/K21 Bias Network Characteristics
Characteristics
Parameter 11612V Option K11 11612V Option K21
 Operation
 Dynamic Range
Network Analyzer System Performance
Measurement Uncertainty
 Network Analyzer System Performance
Maximum Output Power
Dynamic Range
 Measurement Port Characteristics
Measurement Uncertainty
 Transmission Measurements
 Index
 Procedure
Index
 Procedure, manual
Description
 148