Specifications

Environmental Requirements

Environmental Requirements

Table 2-1 Environmental Requirements

Parameter

Limits

 

 

Operating temperaturea

+18 °C to +28 °C (+64 °F to +82 °F)

Error-corrected temperature rangeb

±1 °C of measurement calibration temperature

Storage temperature

−40 °C to +75 °C (−40 °F to +167 °F)

 

 

Altitude

 

Operation

< 4,500 meters (≈15,000 feet)

Storage

< 15,000 meters (≈50,000 feet)

 

 

Relative humidity

Always non-condensing

Operation

0 to 80% (26 °C maximum dry bulb)

Storage

0 to 95%

 

 

a.The temperature range over which the calibration standards maintain conformance to their specifications.

b.The allowable network analyzer ambient temperature drift during measurement calibration and during measurements when the network analyzer error correction is turned on. Also, the range over which the network analyzer maintains its specified performance while correction is turned on.

Temperature—What to Watch Out For

Changes in temperature can affect electrical characteristics. Therefore, the operating temperature is a critical factor in performance. During a measurement calibration, the temperature of the calibration devices must be stable and within the range specified in Table 2-1.

IMPORTANT Avoid unnecessary handling of the devices during calibration because your fingers are a heat source.

2-2

85036B/E

Page 14
Image 14