Specifications

Electrical Specifications

Electrical Specifications

The electrical specifications in Table 2-4apply to the devices in your calibration kit when connected with an Agilent precision interface.

Table 2-4 Electrical Specifications

Device

Specification

Frequency (GHz)

 

 

 

Loadsa

Return loss ³ 46 dB (r £ 0.00501)

DC to £ 2

 

Return loss ³ 40 dB (r £ 0.01000)

> 2 to £ 3

 

 

 

a.23° +5 °C; typical resistance change: ±300 ppm/°C

Certification

Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agilent further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (NIST) to the extent allowed by the institute’s calibration facility, and to the calibration facilities of other International Standards Organization members. See “How Agilent Verifies the Devices in This Kit” on page 4-2for more information.

2-4

85036B/E

Page 16
Image 16
Agilent Technologies 85036B manual Electrical Specifications, Certification