1-2 85036B/E
General Information
Calibration Kit Overview

Calibr ati on Kit Overview

The Agilent 85036B and 85036E type-N calibration kits are used to calibrate Agilent
network a nalyzers up to 3 GHz for m easurements of components with 75type-N
connectors.

Kit Contents

The 85036B calibration kit contains the following:
one male and one female 7 5type-N open termination
one male and one female 7 5type-N short termination
one male and one female 7 5type-N load
three 75type-N to 75type-N adapte rs
Refer to Table 6-1 and Fig u re 6-1 for a com plete list of kit contents and their associated
part numbe rs.
The 85036E calibration kit contai ns the following :
one male com bination open/short termination
one male 75 type-N load
Refer to Table 6-2 and Fig u re 6-2 f or a complete list of kit contents and their a ssociated
part numbe rs.
Broadband Loads
The bro adband loads a re inst rument-grade , 75terminations that have been optimized for
perfor mance up to 3 GHz. The rugged internal struc ture provides for highl y repeata ble
connections. A distributed resistive element on sapphire provides ex cellent stabili ty and
return loss.
Opens and Shorts
The opens and shorts are built from parts that are mac hined to the current state-of the- art
pre cision mach inin g.
The short’s inner conductors have a one- piece construction, c ommon with the shorting
plane . Thi s constructi on provides for extremely repeatable connections.
The female open has a s eparate-piece inner conductor that is ma de from a
low-dielectric-c onstant plastic to minimize com pensation v alues.
Both the opens and shorts are constructed so that the pin depth c an be controlled very
tightly, thereby minimizing pha se errors. Some of the opens a nd shorts have offsets. The
lengths of these offsets ar e designed s o that the difference in phase of their reflecti on
coefficients is approximately 180 d egrees at a ll fr equencies.