Agilent Technologies E, 85036B manual Table A-1 Standard Class Assignments

Models: 85036B E

1 49
Download 49 pages 42.54 Kb
Page 42
Image 42

Standard Definitions

Standard Class Assignments

Standard Class Assignments

Class assignment organizes calibration standards into a format compatible with the error models used in the measurement calibration. A class or group of classes corresponds to the systematic errors to be removed from the measured network analyzer response. Table A-1lists the classes of the devices in this calibration kit.

Table A-1 Standard Class Assignments

Calibration Kit Label: N 75 Ω

Class

A

B

C

D

E

F

G

Standard Class Label

 

 

 

 

 

 

 

 

 

S11A

2

8

 

 

 

 

 

Opens

 

 

 

 

 

 

 

 

 

S11B

1

7

 

 

 

 

 

Shorts

 

 

 

 

 

 

 

 

 

S11C

3

 

 

 

 

 

 

Load

 

 

 

 

 

 

 

 

 

S22A

2

8

 

 

 

 

 

Opens

 

 

 

 

 

 

 

 

 

S22B

1

7

 

 

 

 

 

Shorts

 

 

 

 

 

 

 

 

 

S22C

3

 

 

 

 

 

 

Load

 

 

 

 

 

 

 

 

 

Forward Transmission

4

 

 

 

 

 

 

Thru

 

 

 

 

 

 

 

 

 

Reverse Transmission

4

 

 

 

 

 

 

Thru

 

 

 

 

 

 

 

 

 

Forward Match

4

 

 

 

 

 

 

Thru

 

 

 

 

 

 

 

 

 

Reverse Match

4

 

 

 

 

 

 

Thru

 

 

 

 

 

 

 

 

 

Response

1

7

2

8

4

 

 

Response

 

 

 

 

 

 

 

 

 

Response and Isolation

1

7

2

8

4

 

 

Response & Isolation

 

 

 

 

 

 

 

 

 

A-2

85036B/E

Page 42
Image 42