Performance Verification

The procedures in this section are used to test the switch module’s electrical performance using the specifications in Appendix A of the E1364A User’s Manual as the performance standards. These tests are suitable for incoming inspection, troubleshooting, and preventive maintenance.

Test Fixture A test fixture is required for the following tests. It is recommended that you order an extra terminal module (aka "terminal block") to use as a test fixture, so that you do not have to re-wire the terminal module each time these tests are performed. The terminal module part number is E1364-80001.

Figures 2-1(a) and 2-1(b) show how the test fixture should be wired.

Perform the following steps to wire the test fixture:

Short all NO (Normally Open) lines together.

Short all NC (Normally Closed) lines together.

Short all C (Common) lines together.

Figure 2-1(a). Agilent E1364A Test Fixture Schematic

Chapter 2

Verification Tests 21

Page 21
Image 21
Agilent Technologies service manual Performance Verification, 1a. Agilent E1364A Test Fixture Schematic