Agilent Technologies E1364A service manual Normally Closed to

Models: E1364A

1 60
Download 60 pages 5.54 Kb
Page 28
Image 28

Normally Closed to

Common 1. Hardware Connections

Figure 2-5. DC Isolation Test (NC Channels to Common)

2.Equipment Setup

Set DMM to: 2-wire ohms, 1 Grange.

Send *RST to the switch.

Send the CLOS (@100:115) command to the switch to open all NC contacts.

3.DC Isolation Reading

Trigger the DMM and record the reading.

28 Verification Tests

Chapter 2

Page 28
Image 28