Agilent Technologies E1364A service manual Test 2-1 Closed-Channel Resistance Test

Models: E1364A

1 60
Download 60 pages 5.54 Kb
Page 23
Image 23

Test 2-1: Closed-Channel Resistance Test

The purpose of this test is to verify that all relay contacts meet the closed-channel resistance specification for the switch module. If the closed-channel resistance of any contact is greater than 3.5 , the relay should be replaced.

Normally Open

Contacts 1. Hardware Connections

Figure 2-2. Closed-Channel Resistance (NO channels)

2.Equipment Setup

Set DMM to: 4-wire ohms, autorange.

Send *RST to the switch to open all NO contacts.

3.Closed-Channel Reading (channel 00)

Send CLOS (@100) to the switch to close channel 00.

Trigger the DMM and record the reading.

4.Open-Channel Reading (channel 00)

Send OPEN (@100) to the switch to open channel 00.

Trigger the DMM and verify that an open circuit is indicated (>108 ).

5.Closed- and Open-Channel Readings (channels 01-15)

Repeat steps 3 and 4 for channels 01 through 15. Use

CLOS (@ccnn) and OPEN (@ccnn), where

cc= switch card number (01-99, leading zero not necessary) nn = channel number (00-15)