Agilent Technologies E1364A service manual DC Isolation Test NC Channels to Chassis

Models: E1364A

1 60
Download 60 pages 5.54 Kb
Page 30
Image 30

Normally Closed to

Chassis 1. Hardware Connections

Figure 2-7. DC Isolation Test (NC Channels to Chassis)

2.Equipment Setup

Set DMM to: 2-wire ohms, 1 Grange.

Send *RST to the switch to close all NC contacts.

3.DC Isolation Reading

Trigger the DMM and record the reading.

30 Verification Tests

Chapter 2

Page 30
Image 30
Agilent Technologies E1364A service manual DC Isolation Test NC Channels to Chassis