122 Error Messages
Table D-1. Summary of System Error Messages (continued)
Error
Number Error String [Description/Explanation/Examples]
-141 Invalid character data [bad character, or unrecognized]
-144 Character data too long [maximum length is 12 characters]
-148 Character data not allowed [character data not accepted where positioned]
-150 String data error [generic string error]
-151 Invalid string data [e.g., END received before close quote]
-158 String data not allowed [string data not accepted where positioned]
-160 Block data error [generic data block error]
-161 Invalid block data [e.g., END received before length satisfied]
-168 B lo ck data not allo wed [block data not accepted where positioned]
-220 Parameter error
-221 Settings conflict [uncoupled parameters]
-222 Data o ut of range [e.g., outside the range of this device]
-223 Too much data [out of memory; block, string, or expression too long]
-240 Hardware error [device-dependent]
-241 Hardware missing [device-dependent]
-310 System error [device-dependent]
-313 Calib ration memory lost [out of calibration due to memory failure]
-330 Self-test failed [more specific data after “;”]
-350 Queue overflow [errors lost due to too many errors in queue]
-400 Query error [generic query error]
-410 Query INTERRUPTED [query followed by DAB or GET before response complete]
-420 Query UNTERMINATED [addressed to talk, incomplete programming message received]
-430 Query DEADLOCKED [too many queries in command string]
-440 Query UNTERMINATED [query received after query for indefinite response]
0 No error
1 CAL jumper prevents calibration
2 CAL passcode is incorrect
3 CAL not enabled
4 Computed readback CAL constants are incorrect
5 Computed programming CAL constants are incorrect
6 Wrong sequence of calibration commands
7 CV or CC status is incorrect for this command
100 Table data not monotonic
101 Tables not same length [# of voltage points different from # of current points]
102 Table full
103 No table selected
104 Table in use
105 Too many tables
106 Empty table
107 Table has points exceeding maximum output power