66    Language Dictionary
 Command Syntax *RCL <NRf>
 Parameters 0 | 1 | 2 | 3
 Example *RCL 3
 Query Syntax (None)
 Related Commands *PSC   *RST   *SAV
 *RST
 Meaning and Type
 Reset    Device State
 Description
 This command resets the Agilent SAS to a factory-defined state as defined below. *RST also forces an ABORt command.
If Simulator  or Table mode had previously been programmed, the operation of the unit returns to Fixed mode. Simulator
mode settings revert to the factory default values, and no tables are selected.
  Command   State
  CAL:STAT OFF   DISP[:WIND]:MODE NORM
  CURR[:LEV][:IMM] *   DISP[:WIND]:TEXT
  CURR[:LEV]:TRIG *   INIT:CONT OFF
  CURR:MODE  FIXed   OUTP[:STAT] OFF
  CURR:PROT MAX   OUTP:PROT:DEL *
  CURR:PROT[:LEV] *   TRIG:SOUR BUS
  CURR:PROT:STAT OFF   VOLT[:LEV][:IMM] *
  CURR:SAS:ISC IMAX   VOLT[:LEV][:TRIG] *
  CURR:SAS:IMP 0.8 X IMAX   VOLT:PROT[:LEV] *
  CURR:TABL:OFFS 0   VOLT:SAS:VOC VMAX
  DIG:DATA 0  VOLT:SAS:VMP 0.8 X VMAX
  DISP[:WIND]:STAT ON   VOLT:TABL:OFFS 0
      * Model-dependent value. See Table 7-3.
 Command Syntax *RST
 Parameters (None)
 Query Syntax (None)
 Related Commands *PSC *SAV
 *SAV
 Meaning and Type
 SAVE    Device State
 Description
 This command only applies in Fixed operating mode. It saves the present state of the Agilent SAS to the specified location
in non-volatile memory.  Up to four states can be stored.  Under certain conditions (see Turn-On Conditions in chapter 5),
location 0 may hold the device state that is automatically recalled at power turn-on. Simulator and Table mode states cannot
be saved by this command. The following Agilent SAS parameters are stored by *SAV:
CURR[:LEV][:IMM] OUTP[:STAT] OUTP:REL:POL
CURR:PROT:STAT OUTP:PROT:DEL VOLT[:LEV][:IMM]
DIG:DATA[:VAL] OUTP:REL[:STAT] VOLT:PROT[:LEV]