88    Status Reporting
 Table 8-2.  Bit Configurations of Status Registers
 Bit  Signal  Meaning   Bit  Signal  Meaning
   Operation Status Group      Standard Event Sta tus Group
 0 CAL  The interface is computing new
calibration constants.   0 OPC  Operation complete.
 5 WTG  The interface is waiting for a trigger.   2 QYE  Query error.
 8 CV  The power module is in constant
voltage mode.   3 DDE  Device-dependent error.
 10  CC  The power module is in constant   4 EXE  Execution error.
    current mode.   5 CME  Command error.
       7 PON  Power on.
   Questionable Status Group      Status Byte and Service Request
Enable Registers
 0 OV  The power module overvoltage
protection circuit has tripped.   3 QUES  Questionable status summary bit.
 1 OC  The power module overcurrent
protection circuit has tripped.   4 MAV  Message Available summary bit.
 4 OT  The power module has an
overtemperature condition.   5 ESB  Event Status summary bit.
 9 RI  The power module remote inhibit state
is active.   6 MSS
 RQS  Master Status summary bit.
 Request Service bit.
 10  UNR  The power module output is
unregulated.   7 OPER  Operation status summary bit.
  Figure 8-1. Agilent SAS Status Model