Language Dictionary    75
 OUTP:PROT:DEL
 This command only applies in Fixed mode. It sets the time in seconds between the programming of an output change that
produces a CV, CC, or UNREG condition and the recording of that condition by the Status Operation Condition register.
The delay prevents the momentary changes in Agilent SAS status that can occur during reprogramming from being
registered as events by the status subsystem.  Since the delay applies to CC status, it also delays the OCP (Fixed mode
overcurrent protection) feature.  The OVP (overvoltage protection) and the OC (overcurrent level) is not affected by this
delay.
 Command Syntax OUTPut:PROTection:DELay <NRf>
 Parameters 0 to 32,767 | MIN | MAX
 Units seconds
 *RST Value 0
 Examples OUTPUT:PROTECTION:DELAY 75E-1
 Query Syntax OUTPut:PROTection:DELay?
 OUTPut:PROTection:DELay? MIN
 OUTPut:PROTection:DELay?  MAX
 Returned Parameters <NR3>    OUTP:PROT:DEL? returns value of programmed delay. 
OUTP:PROT:DEL?  MIN and OUTP:PROT:DEL?  MAX return 
the minimum and max imum programmab le delays.
 Related Commands *RST    *RCL    *SAV    CURR:PROT:STAT
 [SOUR:]CURR
 [SOUR:]CURR:TRIG
 These commands only apply in Fixed mode. They set the immediate current level or the pending triggered current level of
the Agilent SAS.  The immediate level is the current programmed for the output terminals.  The pending triggered level is a
stored current value that is transferred to the output terminals when a trigger occurs.  A pending triggered level is unaffected
by subsequent CURR commands and remains in effect until the trigger subsystem receives a trigger or an ABORt
command is given.  If there is no pending CURR:TRIG level, then the query form returns the CURR level.  In order for
CURR:TRIG to be executed, the trigger subsystem must be initiated (see INITiate).
 Command Syntax [SOURce]:CURRent[:LEVel] [:IMMediate][:AMPLitude] <NRf+> 
[SOURce][:CURRent[:LEVel]:TRIGgered [:AMPLitude] <NRf+>
 Parameters Table 7-3
 Default Suffix A
 *RST Value Table 7-3
 Examples CURR 200 MA        CURRENT:LEVEL 200 MA
 CURR:TRIG 20         CURRENT:LEVEL:TRIGGERED 20
 Query Syntax [SOURce]:CURRent[:LEVel] [:IMMediate][:AMPLitude]? 
[SOURce]:CURRent[:LEVel] [:IMMediate][:AMPLitude]?  MAX 
[SOURce]:CURRent[:LEVel] [:IMMediate][:AMPLitude]?  MIN 
[SOURce]:CURRent[:LEVel]:TRIGgered [:AMPLitude]? 
[SOURce]:CURRent[:LEVel]:TRIGgered [:AMPLitude]?  MAX 
[SOURce]:CURRent[:LEVel]:TRIGgered [:AMPLitude]?  MIN
 Returned Parameters <NR3> CURR? and CURR:TRIG? return presently programmed immediate and 
triggered levels.  If not triggered level is programmed, both returned values are the same.
 CURR?  MAX and CURR?  MIN return the max imum and minimum pr ogrammable 
immediate current levels.
 CURR:TRIG?  MAX and CURR:TRIG?  MIN return the maxi mum and minimum 
programmable triggered current levels.
 Related Commands   For  CURR       *SAV      *RCL        *RST
 For CURR:TRIG        ABOR      CURR     *RST