Remote Programming    55
The following example illustrates auto-parallel operation in Table mode.
 1000 ! 2 Units in auto-parallel - Table Mode  
 1010 OUTPUT 705;”*RST”  Resets the  master unit.
 1020 OUTPUT 706;”*RST”  Resets the  slave  unit.
 1030 OUTPUT 705;”MEM:TABL:SEL TABLE1”  Selects a table  for entering data for master unit
 1040 OUTPUT 705;”MEM:TABL:VOLT 0, 5, 10, 50, 55, 60”  Enter voltage data for master unit
 1050 OUTPUT 705;”MEM:TABL:CURR 4, 4, 3.5, 3, 2.5, 0”  Enter current data for master unit
 1060 OUTPUT 705;”CURR:TABL:NAME TABLE1”  Activates the table for master unit
 1070 OUTPUT 706;”MEM:TABL:SEL TABLE1”  Selects a table  for entering data for  slave unit
 1080 OUTPUT 706;”MEM:TABL:VOLT 0, 5, 10, 50, 55, 60, 999”  Enter voltage data for slave unit
 1090 OUTPUT 706;”MEM:TABL:CURR 4, 4, 3.5, 3, 2.5, 0, -999”  Enter current data for slave unit
 1060 OUTPUT 706;”CURR:TABL:NAME TABLE1”  Activates the table for slave unit
 1100 OUTPUT 705;”CURR:MODE:TABL;:OUTP:STAT ON”  Enables the output for master unit.
 1110 OUTPUT 706;”CURR:MODE:TABL;:OUTP:STAT ON”  Enables the output for slave unit.
Changing Outputs by Trigger (applies in Fixed mode only)
 If you do not program pending triggered levels, they default to the programmed (immediate) output levels.  The following
statements shows some basic trigger commands.
 OUTP OFF  Disable the output.
 VOLT:LEV:IMM 55;TRIG 60  Program the voltage to 55V and the triggered level to 60 V.
 CURR:LEV:IMM 2.5;TRIG 3  Program the current  to 2.5 A and the triggered level to 3 A.
 VOLT:LEV:IMM? ;TRIG?;:CURR:LEV:IMM?;TRIG?  Check all the programmed values.
 OUTP ON  Enable the output.
 MEAS:VOLT?;CURR?  Read back the immediate levels from the sense terminals.
 INIT;TRIG  Arm the trigger circuit and send a single trigger.
 INIT;*TRG  Same as above, except using a common command.
 MEAS:VOLT?;CURR?  Read back the triggered levels from the sense terminals.
 If you need to send two or  more triggers, program the trigger circuit for continuous arming.
 OUTP OFF  Disable the output.
 VOLT:LEV: IMM 6O;TRIG 55  Program the voltage to 60 V and the triggered level to 55 V.
 INIT:CONT ON  Program the trigger circuit for continuous arming.
 OUTP ON  Enable the output to 60 V.
 TRIG  Trigger the output voltage to 55 V.
 VOLT:TRIG 60;:TRIG  Set the pending trigger level to 60 V and send a single trigger.
 INIT:CONT OFF  Remove the continuous trigger arming.
Saving and Recalling States
 You can remotely save and recall operating states.  See *SAV and *RCL in chapter 7 for more information.
Note: When you turn the Agilent SAS on, it automatically retrieves the state stored in location 0. When a unit is shipped,
this location contains the factory defaults (see *RST in chapter 7).
 OUTP OFF;VOLT:LEV 6.5;PROT 6.8  Program a desired operating state.
 CURR:LEV 335;PROT:STAT ON  
 *SAV 2  Save this state to location  2 .
 *RCL 2  (Later) recall this same state.