Installation and Setup
Table 1-1
NOTE
WARNING
CAUTION
Installation and Setup
Power Requirements
Power Requirements
Description |
| Specification1 |
Voltage, Frequency | 100 to 120 Vrms, 50/60/400 Hz | |
| 220 to 240 Vrms, 50/60 Hz | |
|
|
|
Power Consumption, On | Base | Fully Loaded |
| < 260 W | < 450 W |
|
|
|
Power Consumption, Standby | < 20 W |
|
|
|
|
1. For greater detail, refer to the Specifications Guide.
AC Power Cord
The analyzer is equipped with a
Various AC power cables are available that are unique to specific geographic areas. You can order additional AC power cables for use in different areas. AC Power Cords, on page 11 lists the available AC power cables, illustrates the plug configurations, and identifies the geographic area in which each cable is appropriate.
The front panel switch is a standby switch only; it is not a LINE switch (power disconnecting device).
Install the product so that the detachable power cord is readily identifiable and easily reached by the operator. The detachable power cord is the product disconnecting device. It disconnects the mains circuits from the mains supply before other parts of the product. The front panel switch is only a standby switch and is not a LINE switch. Alternatively, an externally installed switch or circuit breaker (which is readily identifiable and is easily reached by the operator) may be used as a disconnecting device.
Always use the
10 | Chapter 1 |