2 Features and Functions

Function Limitations

If you change to a function where the maximum frequency is less than the current function, the frequency will be adjusted to the maximum value for the new function.

For example, if you are currently outputting a 20 MHz sine wave and then change to the Ramp function, the U2761A will automatically adjusts the output frequency to 200 kHz (the upper limit for Ramp).

Duty Cycle Limitations

For Square waveforms, the U2761A may not be able to use the full range of duty cycle values at higher frequencies as shown below.

20% to 80% (frequency 10 MHz)

40% to 60% (frequency > 10 MHz)

If you change to a frequency that cannot produce the current duty cycle, the duty cycle is automatically adjusted to the maximum value for the new frequency.

For example, if you currently have the duty cycle set to 70% and then change the frequency to 12 MHz, the U2761A will automatically adjusts the duty cycle to 60% (the upper limit for this frequency).

Soft Front Panel Operation

On the Frequency panel as shown in Figure 2- 3, input the desired frequency value and select the unit from the drop down list. To set the waveform period instead, select Period .

Figure 2-3Panel view of the frequency section

30

U2761A User’s Guide

Page 48
Image 48
Agilent Technologies U2761A manual Function Limitations, Duty Cycle Limitations