Agilent Technologies U2761A manual Generate AM Using Scpi Commands

Models: U2761A

1 123
Download 123 pages 50.59 Kb
Page 72
Image 72

2 Features and Functions

Generate AM Using SCPI Commands

Example 1

 

-> *CLS; *RST

//To reset the U2761A to default

 

power-on state, this command

 

can be ignored if this operation

 

is not required.

-> AM:STAT ON

//Enables AM.

-> FUNC SQU

//Sets the carrier waveform to

 

Square wave.

-> FREQ 2000

//Sets the carrier frequency to

 

2 kHz.

-> VOLT 5 VPP

//Sets the output amplitude to

 

5 Vpp.

-> VOLT:OFFS 0

//Sets the output offset to 0.

-> AM:INT:FUNC SIN

//Changes the modulating

 

waveform shape to Sine wave.

-> AM:INT:FREQ 500

//Changes the modulating

 

frequency to 500 Hz.

-> AM:DEPT 50

//Changes the modulation depth

 

to 50%.

-> OUTP ON

//Turns on output.

54

U2761A User’s Guide

Page 72
Image 72
Agilent Technologies U2761A manual Generate AM Using Scpi Commands