Chapter 1 General Information
Preventive Maintenance
Site Master preventive maintenance consists of cleaning the unit and inspecting and clean- ing the RF connectors on the instrument and all accessories.
Clean the Site Master with a soft,
CAUTION: To avoid damaging the display or case, do not use solvents or abra- sive cleaners.
Clean the RF connectors and center pins with a cotton swab dampened with denatured alco- hol. Visually inspect the connectors. The fingers of the N (f) connectors and the pins of the N (m) connectors should be unbroken and uniform in appearance. If you are unsure whether the connectors are good, gauge the connectors to confirm that the dimensions are correct.
Visually inspect the test port cable(s). The test port cable should be uniform in appearance, not stretched, kinked, dented, or broken.
Calibration
The Site Master is a field portable unit operating in the rigors of the test environment. An
NOTES:
For best calibration
For best results, use a phase stable Test Port Extension Cable (see Optional Accessories). If you use a typical laboratory cable to extend the Site Master test port to the device under test, cable bending subsequent to the OSL calibration will cause uncompensated phase reflections inside the cable. Thus, cables which are NOT phase stable may cause measurement errors that are more pro- nounced as the test frequency increases.
For optimum calibration, Anritsu recommends using precision calibration com- ponents.
1 - 6