![](/images/new-backgrounds/1000866/866131x1.webp)
Chapter 5
Spectrum Analyzer
Measurements
Introduction
This chapter provides a description of field measurements available when the Model S114C or S332C Site Master is in Spectrum Analyzer mode.
Occupied Bandwidth
A common measurement performed on radio transmitters is that of occupied bandwidth (OBW). This measurement calculates the bandwidth containing the total integrated power occupied in a given signal bandwidth. There are two different methods of calculation de- pending on the technique to modulate the carrier.
99% Method — The occupied frequency bandwidth is calculated as the bandwidth containing 99% of the transmitted power.
X dB Down Method — The occupied frequency bandwidth is defined as the band- width between the upper and lower frequency points at which the signal level is X dB below the peak carrier level.
Required Equipment
Site Master Model S114C or S332C
30 dB, 50 Watt,
Test Port extension cable, Anritsu 15NNF50 – 1.5C
Procedure |
| 5 |
| |
Step 1. Using the test port extension cable and 30 dB attenuator connect the Site Master |
|
| ||
| to the RF_Out test port. |
|
| |
Step 2. | Press the ON/OFF key on the Site Master. |
|
| |
Step 3. Press the MODE key. |
|
|
| |
|
|
| ||
Step 4. Use the Up/Down arrow key to select the Spectrum Analyzer mode and press |
|
| ||
| ENTER. |
|
| |
Step 5. Enter the center frequency of the measurement. |
|
| ||
Step 6. Select the appropriate frequency span, reference level, input attenuation level, |
|
| ||
| resolution bandwidth, and video bandwidth. |
|
| |
Step 7. | Press the MEAS key. |
|
| |
Step 8. | Press the OBW soft key. |
|
| |
Step 9. | Select measurement method, % or dBc down by pressing the corresponding soft |
|
| |
| key and press ENTER. |
|
|