Chapter 5 Spectrum Analyzer Measurements
Step 10. Press the MEAS key to initiate actual measurement. The Site Master will display the measurement results in the display area.
The Site Master default value for the % method is 99%, and N dBc down is 3 dB down.
The table below provides typical instrument settings for measuring a CW and an
Table
| CW Wave |
|
|
|
|
Frequency Span | 900 MHz | 900 MHz |
|
|
|
Reference Level | ||
|
|
|
Input Attenuation Level | Auto | Auto |
|
|
|
Resolution Bandwidth | Auto | 10 kHz |
|
|
|
Video Bandwidth | Auto | 30 kHz |
|
|
|
Channel Power Measurement
One of the most common measurements for a transmitter is that of channel power. This test measures the output power, or channel power, of a transmitter over a frequency range and a specific time interval. Channel Power measurements are used to validate transmitter perfor- mance, verify compliance with FCC regulations, and keep overall system interference at a minimum.
Channel Power Measurement with the Site Master
The Site Master can be used to measure CDMA signals in terms of channel power. It is nec- essary to use precise and specific settings to accurately measure the CDMA signal because the CDMA is different from CW signals.
For accurate
CDMA Channel Bandwidth: 1.23 MHz
RBW setting: 10 kHz
VBW 3 RBW or at least 30 kHz
Averaging function turned off
Required Equipment
Site Master Model S114C or S332C
30 dB, 50 Watt,
Test Port extension cable, Anritsu 15NNF50 – 1.5C