Daily Operations
Before each day's use of the Jerome
•Press the power ON button.
The digital meter displays 000.
(Disregard the digital meter's initial momentary reading.)
Recharge or replace the battery pack if the LO BAT indicator REMAINS ON. Refer to “Charging Batteries” on page 17 and/or “Replacing the Battery Pack” on page 21.
To ensure the instrument's electronics have stabilized, allow a
•Use the Zero Air Filter to equilibrate the instrument to ambient air temperature. Install the Zero Air Filter in the instrument’s intake.
Sample continuously until the readings stabilize.
•Perform sensor regeneration. Refer to page 12 for the procedure.
•Thirty minutes after sensor regeneration is complete, zero the instrument. See page 13.
NOTE: For maximum accuracy, such as when testing with the Functional Test Module, wait 30 minutes after the sensor regeneration cycle to
• Press the SAMPLE button.
During the sample cycle, the digital meter displays bars
•At the end of the sampling cycle, read the digital
meter.
The number shown on the digital meter is the hydrogen sulfide concentration in ppm.
This value remains on the display until the next sample is taken.
The digital meter automatically zeroes at the start of each sample.
• At the end of each day's use, perform sensor regeneration as described in the next section.
DO NOT ALLOW HYDROGEN SULFIDE TO STAY
ON THE GOLD FILM SENSOR OVERNIGHT.