5. MAINTENANCE

Preventive Maintenance Calendar

To keep the Jerome 631-X operating at peak performance, follow the maintenance schedule below as a guide. Since maintenance is more a function of application and amount of use rather than time, your requirements may be different from the listed schedule. Call AZI Customer Service at 800-528-7411, 602-470-1414, or e-mail support@azic.com for additional guidance for your environment and operation.

 

PART/COMPONENT

 

 

MAINTENANCE CYCLE

 

 

REFER TO

 

 

 

 

 

 

PAGE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Charge batteries

 

At least once per month, after 1 month's storage,

 

 

Page 17

 

 

or when LO BAT appears

 

 

 

 

 

 

 

 

 

 

Change .25 inch fritware

 

 

Weekly or as needed

 

 

Page 19

 

Change internal filters and

 

After 6 months of use or as needed

 

 

Page 20

 

tubing

 

 

 

 

 

 

 

 

 

 

 

Replace zero air filter1

 

Annually

 

 

 

 

Factory calibration

 

 

Annually

 

 

Page 23

 

Calibration check

 

 

Monthly or as needed

 

 

Appendix

 

 

 

 

 

A, Page 36

 

 

 

 

 

 

 

 

 

 

 

 

Annually or as needed.

 

 

Page 21

 

Replace batteries

 

The battery pack contains NiCad batteries.

 

 

 

 

 

 

 

Dispose of properly.

 

 

 

 

NOTE: Install the zero air filter into the instrument's intake during storage.

1Zero air filters, LFS and LFD scrubber filters contain Resisorb™. Call AZI Customer Service at 800-528-7411, 601-470-1414, or e-mail support@azic.com for a copy of the Material Safety Data Sheet. Dispose of all filters properly.

Page 18 of 50

AZI Customer Service 800-528-7411 or 602-470-1414

Page 18