Arizona Jerome 631-X operation manual Instrument Operation, LCD Codes

Models: Jerome 631-X

1 50
Download 50 pages 58.46 Kb
Page 10
Image 10

4. INSTRUMENT OPERATION

LCD Codes

 

 

 

 

 

LCD CODE

EXPLANATION

 

 

000

Ready to sample

 

 

 

 

 

 

.000

No hydrogen sulfide reading

 

 

 

 

 

 

.8.8.8

Sensor saturated-regeneration needed (refer to page 12)

 

 

 

 

 

 

.H.H.H

Sensor regeneration in progress (.H.H.H flashes)

 

 

.L.L.L

Re-zero needed (refer to page 13)

 

 

 

 

 

 

.P.P.P

Power cord required or low line power, <100 VAC (or 200 VAC)

 

 

 

(see pages 16 and 17, Changing the Fuse, if .P.P.P remains on after

 

 

 

the cord is connected.)

 

 

.LO BAT

Recharge batteries (refer to page 17)

 

 

 

 

 

 

.E.E.E

Same as LO BAT, automatically shuts off

 

 

 

 

 

 

.HL

Very high concentration has been detected. Refer to your safety

 

 

 

policy for additional direction to confirm the concentrations."

 

 

DURING SAMPLING

 

 

 

.-

0-25% sensor saturation

 

 

 

 

 

 

.--

25-50% sensor saturation

 

 

 

 

 

 

.---

50-75% sensor saturation

 

 

-.---

75-100% sensor saturation

 

 

 

 

 

 

DURING SAMPLING, USING SURVEY MODE

 

 

-

Survey sampling (minus sign flashes continuously)

 

WHEN ZERO IS

Adjust to 0 only after sensor regeneration. It is normal for the

DEPRESSED

display to read H after sampling has started.

0

Zero, ready to sample

 

 

H

High, turn Zero potentiometer counterclockwise

 

 

L

Low, turn Zero potentiometer clockwise

Page 10 of 50

AZI Customer Service 800-528-7411 or 602-470-1414

Page 10
Image 10
Arizona Jerome 631-X operation manual Instrument Operation, LCD Codes, Display to read H after sampling has started