Optional Communications Capability

Alarm output

30V DC, 100mA

Data output

1.

RS-232 Serial, Baud Rate 1200 for use with data

 

 

logger, and/or Jerome communication program.

 

2.

RS-232 Serial data format with 0 & 20mA current

 

 

logic levels; Baud Rate 1200 (special industrial

 

 

applications) and Analog 20 mA output.

"OPTION BOARD" - See APPENDIX D - JEROME 631-X OPTION BOARD on page 40.

Analog output

0 to 2V or 4 to 20 mA

Auto sample interval

5, 15, 30, or 60 minutes

 

 

Auto regeneration interval

6, 24 or 72 hours

 

 

Instrument I/O Interface

The 631-X I/O port (25 pin D-sub) provides the following functions:

Serial data communication

Interface type: RS-232C full duplex, DCE

Parameters: 1200 Baud, 1 start bit, 8 data bits, 2 stop bits, no parity

Pin assignments:

 

Pin 1

Protective ground

Pin 2

Data in

Pin 3

Data out

Pin 7

Data ground

Serial current loop

Interface type: 20mA current loop, full duplex

Parameters - 1200 Baud, 1 start bit, 8 data bits, 2 stop bits, no parity

Pin assignments:

Pin 1 Protective ground

Pin 4 Data out (+)

Pin 5 Data in (+)

Pin 14 Data out (-)

Pin 16 Data in (-)