13. APPENDIX C - INTERNAL DIP SWITCH SETTINGS

The DIP switch identified as SW2 on the 631-X control board is located at the top center of the board.

CAUTION:

Switch the instrument’s power to OFF before opening the instrument

and changing DIP switch settings.

DIP Switch Settings:

Switch #

Position

Function

Comment

1

ON

AUTORANGE

DISABLED (SEE DIP SWITCHES 5 & 6)

1

OFF

AUTORANGE

ENABLED

2

X

NOT USED

 

3

X

NOT USED

 

4

ON

ZERO

DISPLAY:

 

00-99

 

 

 

 

 

 

4

OFF

ZERO

DISPLAY:

 

L-O-H

 

 

 

 

 

 

 

Switch #

Function

Comment

5

6

RANGE

SAMPLE TIME

ON

ON

0

(.000

- .099PPM)

30 SECONDS

ON

OFF

1

(0.10

- 0.99 PPM)

25 SECONDS

OFF

ON

2

(01.0

- 09.9 PPM)

16 SECONDS

OFF

OFF

3

(10 - 50 PPM)

13 SECONDS