8. JEROME 631-X TECHNICAL SPECIFICATIONS

Range

0.003ppm (3ppb) to 50ppm H2S in four graduated ranges

Sensitivity

0.003ppm H2S

Precision

5% relative standard deviation

 

 

Accuracy

Range 0: " 0.003ppm at 0.050ppm H2S

 

Range 1: " 0.03ppm at 0.50ppm H2S

 

Range 2: " 0.3ppm at 5.0ppm H2S

 

Range 3: " 2ppm at 25ppm H2S

Response time-sample mode

13 seconds

10 to 50 ppm

1.0 to 10.0 ppm

16 seconds

0.10 to 1.00 ppm

25 seconds

0.001 to 0.100 ppm

30 seconds

Response time-survey mode

3 seconds

0 to 50 ppm

1.0 to 9.9 ppm

6 seconds

0.10 to 0.99 ppm

15 seconds

0.001 to 0.099 ppm

20 seconds

Flow rate

150cc/min (0.15 liters/min)

 

 

Power requirements

100-120 V~, 50/60 Hz, 1 A or 220-240 V~, 50/60 Hz, 1 A

 

 

Fuse

F1A T250V, 5mm X 20mm

 

 

Internal battery pack

Rechargeable Nickel Cadmium

 

 

Operating environment

0E to 40EC, non-condensing, non-explosive

 

 

Case construction

Aluminum alloy

 

 

Dimensions

15 cm x 33 cm x 10 cm (6" w x 13" l x 4" h)

 

 

Weight

3.18 kilos (7 pounds)

 

 

Digital meter display

Liquid crystal display (LCD)

 

 

Certification

CE mark on 220-240 V~, 631-XE model only.

 

 

AZI Customer Service 800-528-7411 or 602-470-1414

Page 27 of 50

Page 27
Image 27