Bacharach 2772-0803 manual 2.7. 0 - 1 VOLT CONVERSION, 1. 0 - 1 Volt Conversion 2.8 SYSTEM CHECK

Models: 2772-0803

1 49
Download 49 pages 53.56 Kb
Page 18
Image 18
2.7.0 - 1 VOLT CONVERSION

REMOTE INTELLIGENT SENSOR - AREA MONITOR

2.6.3REACTIVE GASES. Many gases are extremely liable to attenuation and no extension of input tube should be attempted. These gases include ISOCYANATES, HYDRAZINES and ACID GASES such as HF and HCI.

2.6.4SYSTEM PERFORMANCE WITH EXTENDED SAMPLE LINES. Bacharach will not guarantee system performance and accuracy if extended sample lines are fitted, except where the company has expressly given written approval.

NOTE: Without such specific approval, the user must determine that performance is not adversely affected under the particular application conditions.

WARNING

Extended input lines should never be used when sampling low vapor pres- sure compounds such as MDI. These compounds are present in aerosol form, and sampling efficiency will be drastically reduced if the sample line is extended. Instruments intended for aerosol sampling have a dual input tube approximately 1" (25 mm) long protruding from the bottom of the instrument.

2.7.0 - 1 VOLT CONVERSION

2.7.1TOOLS & MATERIALS REQUIRED

2.5 mm Hex Key

Soldering Iron

Solder

Wire Cutter (Small)

Buss Wire (#22 AWG, 1/2" Long)

2.7.2PROCEDURE. Remove main power from RIS System: then set ON/OFF switch SW1 to OFF. Follow the removal of the mechanical chassis from Section 4.11.2.

Remove link (0 Ohm resistor) from the 4 - 20 OUTPUT position using wire cutters to cut both ends of the link. Solder buss wire link into the 0-1 Volt position per Figure 2-1.

Reinstall the mechanical chassis, cables, and hoses using the first two paragraphs of Section

0-1V

4-20MA

J4

POWER

SW1

J3

ON

1

J2

1

1VR1

 

J5

 

 

 

 

 

 

 

BATTERYPACK

 

 

 

 

 

 

TR1

 

 

 

 

 

 

C1 R1 R2IC1

 

 

R3 R4 R5R6 C3C4

 

J1

 

 

 

 

 

C2

 

 

12VDC

-

1

 

 

 

 

 

 

 

INPUT

+

2

+

 

 

 

 

 

 

 

 

C7

 

L1

R8 R9 R10

IC3

 

 

3

 

 

 

 

 

 

 

 

AOV

4

IC2

 

 

 

 

 

D1

ANALOG

 

 

 

 

 

 

 

OUTPUT

 

 

C5

 

 

 

 

 

 

 

AOP

 

+

 

 

 

 

 

 

 

 

 

 

 

 

 

 

C6

 

AIN

 

 

 

 

 

 

 

 

NETWORK

 

VR2ZERO

 

 

 

 

 

 

COMMS

AOUT

 

 

R12

 

R7

R14

R15

 

INTERFACE

B

 

 

IC4

 

 

 

 

 

 

 

 

 

 

SR11

 

 

PI

 

VR3SPAN

 

Q1

 

 

 

 

 

GMD11760

SERIAL

PO

 

R13

 

 

PRINTER

PRDY

 

 

 

 

INTERFACE

 

 

 

 

 

 

PPRS

 

13

 

 

 

 

OV

 

14

0-1V

4-20MA

 

+

 

 

15

 

 

RLA1

-

R16

RLA2

RLF

 

RLA1

 

+

 

 

 

 

RLF

 

 

 

 

 

 

 

-

 

 

 

 

 

RLA2

+

 

 

 

R17

R18

-

20

Figure 2-1. 0 - 1 Volt Conversion

2.8SYSTEM CHECK

NOTE: This procedure should be followed every time a new cassette is installed, to insure the system is operating correctly. During the initial system check of this unit, and at least on a yearly basis there after, the sample inlet tube should be checked for proper installation.

2.8.1TAPE CASSETTE LOADING

WARNING

This involves opening the RIS door, which automatically initiates the door-open alarm. Before opening the door to load or change a cassette, ensure that any external warning system is disabled or that those involved are aware of your intended actions.

Open the door with the key provided. Leave the system switch on. Press the TEST Mode switch on the bottom left inside corner of the door. Open the tape gate with the lever and remove the old cas- sette, if installed. The cassette and its mounting spigots are designed to fit tightly; use a firm and direct pull to disengage the cassette. See Figure 1-2 for the location of the items referred to.

NOTE: Make sure that the ‘O’ ring installed in the cassette molding, (and which seals purge connections), is not left behind when the cassette is removed.

Page 2-4

Instruction 2772-0803

Page 18
Image 18
Bacharach 2772-0803 2.7. 0 - 1 VOLT CONVERSION, 1. 0 - 1 Volt Conversion 2.8 SYSTEM CHECK, Tools & Materials Required