to begin the infusion. Check for drops in the drip chamber.

Continuous Mode

chamber.

VTBD and TIME

1.Press VTBD.

2.Enter bag volume in ml.

3.Press OK. (To leave screen without saving changes, press EXIT.)

4.Press TIME.

5.Enter infusion time in hours and minutes.

NOTE: For a 2 hour infusion, enter 2, 0, 0. The pump will show: Time: 02h00m

6.Press OK. (To leave screen without saving changes, press EXIT.) The pump calculates the rate and asks: Accept Rate?

7.Press OK.

8. Press

RATE and TIME

1.Enter rate in ml/hr.

2.Press TIME.

3.Enter infusion time in hours and minutes.

4.Press OK. (To leave screen without saving changes, press EXIT.)

 

5. Press

to begin the infusion. Check for drops in the drip

 

 

chamber.

To Change the Rate

1.

Press

to stop the infusion.

2.

Press

. The display will reset to 000.0 ml/h.

3.

Enter a new rate.

4.

Press

to begin the infusion. Check for drops in the drip chamber.

To Change the Rate While Infusing

1.Enter a new rate.

2.Press OK.

NOTE: If the OK key is not pressed within 10 seconds of the change, the infusion will continue at the old rate.