Braun 950787 manual Alarms, Alarm Possible Cause Action

Models: 950787

1 27
Download 27 pages 3.58 Kb
Page 23
Image 23

 

OPTIONS

 

 

 

Alarms

 

 

 

 

 

 

 

 

In alarm situations, press

to silence the alarm. Correct the problem. Resume

 

the infusion by pressing

.

 

 

 

 

 

 

 

 

 

 

Alarm

 

Possible Cause

 

Action

 

Air-In-Line!

Air bubble in air-in-line sensor

Remove air from system

 

 

 

 

 

Check Drop

Solution container empty

Replace solution container

 

Sensor!

Roller clamp closed

Open roller clamp

 

Free Flow!

Set loaded incorrectly

Re-load set

 

Not Enough Drops!

Kinks, closed clamps, or

Check for blockage

 

Too Many Drops!

blockage

 

 

 

 

 

Droplets on wall of drip chamber

Shake drip chamber

 

 

Drop sensor out of position

Reposition drop sensor

 

 

 

 

 

Door Open!

Door opened during infusion

Close door

 

 

 

 

 

Dwnstream

Downstream tubing kinked

Correct kinks

 

Occlusion!

Filter blocked

Correct blocked filter

 

 

Closed clamp

Open clamps

 

 

IV positional or infiltrated

Check IV site

 

 

Catheter or vein too small

Increase occlusion limit

 

 

Occlusion limit too low

 

 

 

 

 

 

 

 

 

Invalid data!

Entered data is out of range

Press

; re-enter data

 

 

 

 

 

Keep Vein Open

Programmed infusion is

Stop infusion

 

Rate: Infusion

complete

 

 

 

 

Complete

 

 

 

 

 

 

 

 

 

 

Low Battery!

30 minutes until battery is fully

Silence alarm and connect

 

 

discharged

pump to AC power

 

 

 

 

 

Plug In Now!

Battery is discharged

Connect pump to AC power

 

 

 

 

 

Pump Stopped!

No key panel activity for 2

Silence alarm; begin infusion

 

 

minutes or Standby time expired

 

 

 

 

 

 

 

 

 

Unit Malfunction!

Internal error

Press

until display reads:

 

 

 

 

“Do not press any key until

 

 

 

 

display is off;” wait until pump

 

 

 

 

turns off, turn device on. If

 

 

 

 

alarm repeats, return for

 

 

 

 

service.

 

 

 

Vistabasic Instructions for Use

 

 

22/25

 

 

950787 Rev H 02/03

 

 

 

 

 

Page 23
Image 23
Braun 950787 manual Alarms, Alarm Possible Cause Action