Braun 950787 manual Specifications

Models: 950787

1 27
Download 27 pages 3.58 Kb
Page 26
Image 26

SPECIFICATIONS

Mechanism

Linear peristaltic volumetric infusion pump

Dimensions WxHxD

5.5 in (14cm) x 9.4 in (24cm) x 7.8 in (20cm)

Weight

6.8 lbs. (3.1 kg)

Power Requirements

100-120 VAC, 50-60 Hz, 13 Watts

Leakage Current

Max 30µA

Battery

7.2V, 1.2 Ah NiCd (rechargeable)

Battery Operating Time

=3.5 hours at 125 mL/hr

Rate Range

0.1 to 800.0 mL/hr in 0.1 mL increments with Vista™

 

tubing

Volume to be Infused

0.1 to 9999.9 mL

Range

 

Occlusion Pressure

Variable Settings Low: 8 PSI High: 17 PSI

KVO Rate

1 mL/hr or selected rate, whichever is lower

Flow Rate Accuracy

± 5% measurement time 8hr, temp. 71.6° F

Alarms

Air-in-Line, Low Battery, Plug In Now, Bag Empty,

 

Downstream Occlusion, Door Open, Pump Stopped,

 

KVO, Unit Malfunction

Air-in-Line Alarm

Air bubble size 0.3mL* or 1.5mL/hr**

 

*Can be set from 0.01 to 0.3 mL in service

 

program

 

**Can be set from 0.5 to 3.5 mL/hr in service

 

program

Memory

User option to retain last Continuous and Piggyback

 

Mode data upon power up. Permanent data

 

retention in Program and Ramp/Taper Modes after

 

pump is turned off

Tubing

B. Braun Vista™ administration sets

Modes/Features

Continuous and Piggyback Modes, Program Mode

 

for variable infusions, Ramp/Taper Mode for TPN

 

infusions and Delay A Mode for postponing infusion

 

start and finish

Operating

Operating: 64.4° F to 95° F, 10.15 PSI to 15.37PSI

Temperatures

Non-Operating: -13° F to 131° F, 10.15 PSI to

 

15.37 PSI

Vistabasic Instructions for Use

25/25

950787 Rev H 02/03

 

Page 26
Image 26
Braun 950787 manual Specifications