Braun 950787 manual AC, 12 V, BATTERY, & CC Interface, AC and 12, Battery Operation

Models: 950787

1 27
Download 27 pages 3.58 Kb
Page 24
Image 24

AC, 12 V, BATTERY, & CC INTERFACE

AC and 12 V

Check AC voltage as per rating label.

Insert/screw on power lead (12 V lead in ambulance car) on back.

Battery Operation

Switching to internal battery is automatic in the event of power failure or if 12 V or AC is not connected.

Charge battery on initial use, following a battery alarm, or after 2 months of non-use.

Charging time: approx. 16 hours. Longer charging has no detrimental effect.

Full battery provides operation for 3.5 hours at 125 mL/hr.

Battery capacity is approx. 50% after 2 years.

The service life of the battery can be extended by completely discharging it from time to time and then fully recharging it on AC power.

CC Interface

Interface descriptions available from B. Braun. Connection to interface input uses MFC plug.

Documentation

All operating data of the Vistabasic can be called up and logged via external computer

Vistabasic Instructions for Use

23/25

950787 Rev H 02/03

 